Chassis Disassembly and Reassembly
Service Manual
12-17
Caution:
Ensure the work surface is clean and free of debris.
3.
Orient the oximetry module PCB so the 8-pin J5 connector pins mate to the
parameter board J9 receptacle and the 10-pin J4 connector pins mate to the
parameter board J7 receptacle.
4.
Press downward on the oximetry module PCB until fully seated on the parameter
(interface) board PCB.
5.
Replace and tighten the two T-10 torx screws to a maximum of 6 inch-pounds.
6.
Rotate the parameter board assembly so the oximetry module PCB is facing down-
ward and both connector sockets are facing the front of the bottom enclosure.
7.
Place on the standouts to the left of the bottom enclosure.
8.
Replace and tighten the four T-10 torx screws to a maximum of 10 inch-pounds.
9.
Replace the top cover, ensuring the tabs slip down into the mating slots of the
bottom enclosure.
10.
Replace the faceplate, aligning to the mating tabs of the bottom enclosure and
ensuring the lugs of the top cover slip under the faceplate lip.
11.
Replace the parameter module. Reference
To replace the parameter module
12.
Connect to AC power.
13.
Verify unit functionality. Reference
12.5.2
Monitoring System Chassis Disassembly
Caution:
Observe ESD (electrostatic discharge) precautions when disassembling and
reassembling the monitoring system and when handling any internal
components.
Note:
The battery charge procedure should be performed before monitoring system repairs
when possible.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......