Connection to Nellcor™ Sensors
Service Manual
5-7
5.4
Connection to Nellcor™ Sensors
WARNING:
Use only Covidien-approved sensors and interface cables when connecting to
the sensor port. Connecting any other cable or sensor influences the accuracy
of sensor data, which may lead to adverse results.
The top of the monitoring system screen indicates the sensor type when con-
necting a recommended sensor to the monitoring system or when the moni-
toring system completes POST with an attached sensor.
Note:
Physiological conditions such as excessive patient movement, medical procedures, or
external agents such as dysfunctional hemoglobin, arterial dyes, low perfusion, dark
pigment, and externally applied coloring agents such as nail polish, dye, or pigmented
cream may interfere with the monitoring system’s ability to detect and display
measurements.
Note:
Sensor LED light emissions fall within Class 1 level, according to IEC 60825-1:2001.
To fully connect a Nellcor™ sensor
1.
Firmly connect a Nellcor™ interface cable to the monitoring system’s sensor port.
Reference
, p. 4-3, to identify the port.
2.
Open the plastic latch at the other end of the interface cable.
Figure 5-1.
Sensor Cable insertion into Interface Cable
3.
Plug the interface cable and recommended sensor together.
4.
Snap the plastic latch down over the connectors.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......