Installation
5-2
Service Manual
WARNING:
Disconnect the monitoring system and sensor from the patient during
magnetic resonance imaging (MRI) scanning. Objects containing metal can
become dangerous projectiles when subjected to the strong magnetic fields
created by MRI equipment. Also, induced currents could potentially cause
burns.
WARNING:
To ensure accurate performance and prevent device failure, do not subject
the monitoring system to extreme moisture, such as direct exposure to rain.
Such exposure may cause inaccurate performance or device failure.
WARNING:
Do not connect the monitoring system to an electrical outlet controlled by a
wall switch, since this increases the risk of removal of AC power to the
monitoring system.
WARNING:
Use only Covidien-approved sensors and interface cables when connecting to
the sensor connector. Connecting any other cable or sensor influences the
accuracy of sensor data, which may lead to adverse results.
WARNING:
Use only Covidien-approved interface cables with the monitoring system. Use
of another interface cable will adversely impact performance. Do not attach
any cable intended for computer use to the sensor port.
WARNING:
The monitoring system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, observe the monitoring
system to verify normal operation in the desired configuration.
WARNING:
Ensure the monitoring system is clear of any obstructions that prevent
awareness of visual or audible alarms. Failure to do so may result in
inadvertently missing a visual alarm or an inaudible alarm tone.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......