Operation
6-2
Service Manual
6.2.1
AC Power
When the user connects the monitoring system to an AC power source, if the
internal battery requires charging, the battery condition indicator on the front
panel lights until the internal battery reaches complete charge. In addition,
when the monitoring system is powered on, the battery fuel gauge on the
monitoring screen displays a lightning bolt indicating connection to AC.
If the user powers off the monitoring system while the internal battery is
charging, the battery condition indicator remains lit and the internal fan turns
on until charging completes. Reference
Connection to an AC Power Source
6.2.2
Battery Power
WARNING:
Do not use monitoring system with a depleted battery or in a low voltage
condition.
Battery Status
, p. 5-6, for details on initial internal battery
setup information.
The yellow BATTERY LOW warning flashes and a medium priority alarm sounds
when approximately 14% capacity remains on the existing battery charge. The
red BATTERY CRITICALLY LOW warning flashes and a high priority alarm
sounds when approximately 4% capacity remains on the existing battery
charge. The battery will drain completely and the monitoring system will shut
down if not connected to AC power during a critically low battery condition.
Reference
, p. 6-3, for a description of the low and critical
battery conditions.
To cancel a visual or audible battery condition alarm, connect the monitoring
system to an AC power source. The low battery warning status remains as long
as the battery is in a low voltage condition or until the caregiver presses
DISMISS ALARM for the low battery alarm message.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......