Power Failure Issues
Service Manual
11-11
11.5
Power Failure Issues
Power issues may require replacement of hardware, but may also require iden-
tifying potential issues in the operating environment. Prior to assuming the
problem is with the monitoring system, check the power cable and wall outlet.
Table 11-4.
Power Failure Issues
Problem
Resolution
There is no response when
pressing the POWER ON key.
•
A fuse may be malfunctioning. Replace the fuse.
•
If operating on battery power, the battery may be missing or dis-
charged. If the battery is discharged, charge the battery.
•
If the battery does not charge, replace the battery.
The monitoring system is not
charging the battery even when
connected to an AC power
source.
•
IF the battery is completely depleted and the unit shuts down, it may
not charge after connection to AC and the battery charge indicator
will not light. Power cycle once and retry charging the battery while
connected to AC.
•
Replace the battery.
The monitoring system is oper-
ating on battery power, even
though it is connected to an AC
power source.
•
Ensure the power cord is properly connected to the monitoring
system.
•
Check to see if power is available to other equipment on the same AC
circuit.
Even though it is connected to
an AC power source, the moni-
toring system displays the fol-
lowing System Error when
powered on: “Battery missing
or damaged. Contact a quali-
fied service technician.”
The battery may not be installed. The battery must be installed prior to use.
The monitoring system is not
operating on battery power
when it should.
•
Charge the battery.
•
Replace the battery.
•
Check the battery interconnect board and power cables and replace
as required.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......