6-1
6 Operation
6.1
Overview
This chapter identifies methods for collecting patient oxygen saturation data
while using the Nellcor™ Bedside Respiratory Patient Monitoring System. It
describes menu navigation, power on/off and monitoring screen options,
parameter ranges, Nellcor™ sensor attachments, and configuring default set-
tings suitable for the specific care environment.
Perform regular maintenance and safety checks every 24 months. In the case
of mechanical or functional damage, contact Covidien or a local Covidien rep-
resentative.
6.2
Power
WARNING:
Explosion hazard — Do not use the monitoring system in the presence of
flammable anesthetics.
Caution:
Replace the internal battery every 24 months.
Caution:
Remove and store the internal battery if users expect a significant period of
disuse.
Caution:
A normal power cycle or complete discharge of the battery results in a reset
of all temporary user settings to factory or institutional default settings.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......