![background image](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898087.webp)
Service Mode
Service Manual
6-25
To set the default main trend scale or monitoring history trend scale
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select DISPLAY SETTINGS.
4.
Select the desired option.
a.
MAIN SCREEN TREND SCALE
b.
MONITORING HISTORY TREND SCALE
5.
Select the desired default.
a.
Minutes: 15 or 30
b.
Hours: 1, 2, 4, 8, 12, 24, or 48
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
To adjust default backlight brightness
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select DISPLAY SETTINGS.
4.
Select BACKLIGHT BRIGHTNESS.
5.
Adjust to the desired default brightness.
a.
Press decrement (-) to decrease the default brightness.
b.
Press increment (+) to increase the default brightness.
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
To set backlight OFF permission
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......