Basic Preventive Maintenance
Service Manual
12-9
3.
Clean the surface of the bottom of the monitoring system, using one of the stan-
dard cleaning solutions described for preventive maintenance, to remove any
adhesive residue.
To replace rubber feet
1.
Ensure the bottom surface of the monitoring system is clean and dry.
2.
Peel away the protective release liner on the rubber foot to expose the adhesive.
3.
Gently align the rubber foot to the indented area.
4.
Press down firmly on the rubber foot to properly seat it.
5.
Repeat the process for the remaining rubber feet.
Figure 12-5.
Rubber Feet Replacement
6.
Properly orient and rotate the monitoring system onto its base.
7.
Reconnect to AC power.
8.
Turn on the monitoring system by pressing the ON/STANDBY button.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......