Basic Preventive Maintenance
Service Manual
12-7
12.4.1
Fuse Removal and Replacement
To remove and replace external fuses
1.
Remove the fuse drawer from the power entry module.
a.
Press down on the center tab.
b.
Pull outward.
2.
Remove both 5 x 20-mm, slow blow, 1.5-amp, 250-volt fuses.
Figure 12-3.
External Fuse Removal
3.
Replace with two (2) new, 5 x 20-mm, slow blow, 1.5-amp, 250-volt fuses.
4.
Reinsert the fuse drawer until the tab snaps in place.
12.4.2
Battery or Battery Access Door Replacement
To remove the battery access door and battery
1.
Using the T-15 torx driver, remove and set aside the torx screw from the battery
cover.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......