Power Components Replacement
Service Manual
12-29
Figure 12-16.
Battery Interconnect PCB Removal
To replace the Battery Interconnect PCB
1.
Place the new battery interconnect PCB on the battery cradle.
2.
Replace and tighten both T-10 torx screws to 10 inch-pounds.
3.
Rotate the battery cradle.
4.
Reconnect the five-pin J2 connector.
5.
Replace the cooling fan. Reference
6.
Replace the battery cradle in the back chassis. Reference
Cooling Fan Replacement
To remove the cooling fan
1.
Follow all previous steps for removing the battery cradle. It is not necessary to
remove the power supply PCB or battery interconnect PCB.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......