Modification and Testing
10-16
Service Manual
e.
Select PULSE RATE DELAY.
f.
Ensure the delay is set to OFF.
8.
Turn off the monitoring system by pressing the POWER ON key.
Testing Monitoring Settings
Service technicians may choose to test all monitoring settings in sequence. If
so, reserve turning off the monitoring system until the end of the sequence. If
not, turn it off after completing the individual test to clear all changes.
To test alarm mode
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press MONITORING SETTINGS.
4.
Select ALARM MODE.
5.
Ensure the alarm mode is ADULT ALARM SETTINGS.
6.
Select NEONATE ALARM SETTINGS.
7.
Press SAVE CHANGES.
8.
Ensure the Baby icon appears on the main monitoring screen between both satu-
ration and pulse rate alarm threshold settings.
9.
Press YES.
To test response mode
1.
Turn on the monitoring system by pressing the POWER ON key.
2.
Press MENU.
3.
Press MONITORING SETTINGS.
4.
Select RESPONSE MODE.
5.
Ensure the response mode is NORMAL.
6.
Select FAST.
7.
Press SAVE CHANGES.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......