![background image](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898158.webp)
Modification and Testing
10-36
Service Manual
d.
Pulse waveform peak to peak (P-T-P) amplitude of approximately half-inch.
Actual amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
7.
Press LIGHT LEVEL.
8.
Verify the increase LED lights. The pulse amplitude waveform initially flatlines and
then stabilizes at the same P-T-P amplitude of approximately half-inch.
Note:
Flat-lining is the only indication of a light change at the measurement site. If the
monitoring system recovers and displays normally, this is an indication of proper
operation with light changes.
Figure 10-15.
LIGHT Test: BPM 60, SpO
2
75, MOD low, Light High
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately half-inch P-T-P amplitude. Actual amplitude
may vary but references low pulse amplitude/
high
light indications.
9.
Press PULSE RATE.
10.
Verify the 200 BPM LED lights. The monitoring system registers pulse rate increas-
es and stabilizes to a value between 194 to 206 BPM inclusive.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......