Modification and Testing
10-8
Service Manual
6.
Press EXIT MENU.
7.
Press EXIT AND RESTART.
8.
Press YES.
10.3
Performance Verification
WARNING:
Only qualified service personnel should open the chassis, remove and replace
components, or make adjustments. If your medical facility does not have a
qualified service technician, contact Covidien Technical Services or your local
Covidien representative.
10.3.1
Overview
This section discusses the tests used to verify performance following repairs or
during routine maintenance. All tests can be performed without removing the
monitoring system enclosure. Perform all tests before the battery charge and
battery performance checks, then perform both battery checks as the last oper-
ation before returning the monitoring system to the caregiver. If the monitoring
system fails to perform as specified in any test, make all repairs necessary to
correct the problem before returning the monitoring system to the caregiver.
10.3.2
Required Equipment
Table 10-4.
Equipment and Descriptions
Equipment
Description
Nellcor™ SRC-MAX functional oximetry tester
Provides monitoring system simulation
Nellcor™ Adult SpO
2
Sensor, Reusable
DS-100A
Nellcor™ Adult SpO
2
Nasal Sensor
MAX-A
DOC-10 interface cable
Interface cable
Digital multimeter (DMM)
Fluke Model 87 or equivalent
Safety analyzer
Must meet current AAMI ESI & IEC 60601-1 specifications
Stop watch
Manual or electronic
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......