Power Components Replacement
Service Manual
12-27
Figure 12-15.
Power Supply PCB Removal
1.
Examine the battery cradle for alignment information. Extruded icons on the
battery cradle indicate proper four-pin connector two-pin receptacle orientation.
2.
Orient the new power supply PCB and slide down into the battery cradle retaining
slots and onto the battery cradle standoffs, ensuring the fan cable remains free of
the battery cradle.
3.
Route the fan cable up under the battery cradle so it rests on the appropriate
moulded fin nearest the wire-tied cables.
4.
Replace and tighten both T-10 torx screws to 10 inch-pounds.
5.
Firmly connect the four-pin J2 connector.
6.
Firmly connect the two-pin J1 (AC) and spade E1 (GND) connectors to the recep-
tacles on the opposite side of the power supply board.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......