Repair
12-20
Service Manual
Figure 12-12.
Initial Chassis Disassembly
12.5.3
Monitoring System Chassis Reassembly
To reassemble the monitoring system chassis
1.
Position the front chassis on a static-free work surface, still lying face down, within reach
of the back chassis connectors.
2.
Firmly connect the spade connector (GND) to the J21 mating blade on the Main
PCB.
3.
Firmly connect the five-pin battery connector to the J23 receptacle on the Main PCB.
4.
Firmly connect the four-pin power connector to the J22 receptacle on the Main PCB.
5.
Firmly connect the three-pin fan connector to the J2 receptacle on the Main PCB.
6.
Gently align and place the back chassis onto the front chassis.
Caution:
Do not pinch any wires between the chassis enclosures when
reassembling. Before installing the screws, inspect the entire edge of the
union for even seating.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......