Modification and Testing
10-44
Service Manual
10.8
Monitoring Screen Calibration
Only calibrate the monitoring screen if it stops responding appropriately to human
touch. If caregivers report problems accessing menu items and the screen is
unlocked, consider calibration. Calibration requires a stylus or very fine, rounded
tip of a non-marking utensil. Do not use the pointed tip of a pencil or pen.
To calibrate the monitoring screen
1.
Access the service function menu. Reference
To access service function menus
2.
Select the SERVICE menu.
3.
Select the SERVICE option.
4.
Select TOUCH SCREEN CALIBRATION.
5.
Press YES.
6.
Press NEXT.
7.
Press and hold stylus in exact center of crosshatch in center of monitoring screen.
Figure 10-19.
Initial Calibration Screen
8.
Press and hold stylus in exact center of crosshatch as it moves to each corner of
the monitoring screen.
9.
Insert a USB keyboard in the USB port.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......