444
Appendix C
Error messages
Suffix not allowed
A suffix error. An error other than -131 to -139 occurs in the notation of a suffix.
-138
Suffix not allowed
A suffix is added to a numeric value element to which no suffix should be added.
-134
Suffix too long
The suffix is too long. The suffix contains the notation of 12 characters or more. (Refer to
IEEE488.2,7.7.3.4.)
-102
Syntax error
There is a command or data type that cannot be recognized. For example, although the
4294A does not accept any strings, a string is received.
-310
System error
One of the “system errors” defined for the 4294A occurred.
T
87
THETA MEASUREMENT NOT SELECTED
When the setting of the phase (q) is not active as a measurement parameter, you attempt to
execute a command to set the phase unit (PHAU DEG or PHAU RAD) or a command to set
the phase expansion display (EXPP ON or EXPP OFF) and, as a result, the command is
ignored. The active status of the phase (
q
) setting means:
•
|Z|-
q
or |Y|-
q
is selected as a measurement parameter, and trace B is active.
•
COMPLEX Z-Y is selected as a measurement parameter, and the display format is
POLAR (polar coordinates display format) for the selected active trace.
Before executing a command to set the phase unit or a command to set the phase expansion
display, perform one of the above settings.
-124
Too many digits
The number of digits of the mantissa of the decimal value data element exceeds 255 except
when there are preceding 0s. (Refer to IEEE 488.27.7.2.4.1.)
64
TOO MANY SEGMENTS
You attempt to add segments to the limit line table
at a quantity that exceeds the maximum
number (18). The number of segments you can set in the limit line table is up to 18.
77
TOO MANY SEGMENTS OR POINTS
During the editing of the list sweep table, you attempt to set a value that exceed the
maximum number of segments (18), the maximum number of points per segment (201), or
the maximum number of total points in all segments (801). Set the number of segments or
the number of points to a value that does not exceed the maximum value.
54
TOO MUCH DATA
When data is sent from the controller to the Agilent 4294A in the FORM2, FORM3, or
FORM5 data transfer format, the amount of the sent binary data is too large, or data items
exceed the number of points (NOP) set. Be sure to correctly set the data to be sent or the
number of points of the Agilent 4294A.
-223
Too much data
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......