Chapter 8
279
Analysis and Processing of Result
Save and Recall the Agilent 4294A Internal Data
8. A
nal
ys
is
a
nd P
roc
es
si
ng
of
Res
ult
calibration data and the fixture compensation data.
Table 8-8 shows the contents of the saved calibration data.
Data contents of the trace data (data array, memory array, data trace array, and
memory trace array)
If you specify saving one of the four data items of the trance data (data array, memory
array, data trace array, and memory trace array) on the data save contents selection menu,
the selected data is stored in the data part for each trace (trace A and trace B). The
following example describes the contents of the data part when saving one of the four data
items of the trace data one by one.
Table 8-7
Example of calibration data saved in the data part
"Cal Point: FIXED"
Ü
Ü
"Frequency"_"Comp[open] Real"_"Comp[open] Imag"_"Comp[short] Real"_"Comp[short] Imag"
_"Comp[load] Real"_"Comp[load] Imag"
Ü
4.0000001_2.947073e-09_-1.565653e-09_-6.253917e-03_1.008368e-03_5.01_
-3.311127e-02
Ü
4.5000001_-7.903406e-11_9.789698e-10_-4.783846e-03_-7.028686e-04_5.01_
-4.446183e-02
Ü
5.5000001_-1.688141e_-10-3.870845e-10_-4.990600e-03_-4.369987e-04_5.01_
-4.029654e-02
Ü
..................................................................................
Symbols in the table
Ü
: Linefeed, _ : Tab
Table 8-8
Contents of the calibration data
Setup and data array
Description
"Cal Point: FIXED"
Setting of calibration and compensation
data acquisition points
"Frequency"
Frequency points of calibration and
compensation data
"Comp[open] Real"
OPEN calibration/compensation data real
part
"Comp[open] Imag"
OPEN calibration/compensation data
imaginary part
"Comp[short] Real"
SHORT calibration/compensation data real
part
"Comp[short] Imag"
SHORT calibration/compensation data
imaginary part
"Comp[load] Real"
LOAD calibration/compensation data real
part
"Comp[load] Imag"
LOAD calibration/compensation data
imaginary part
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......