396
Appendix B
Key Definitions
Softkeys displayed by pressing the [Marker
®
] key
MKR ZOOM
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the sweep center value and simultaneously changes the
current sweep span value to the sweep span value specified with the ZOOMING
APERTURE key. In other words, the sweep range is changed so that the marker
position is the new sweep center value and the sweep span is magnified by the
amount specified with the ZOOMING APERTURE key.
MKRZM
ZOOMING APERTURE
Sets the zooming aperture (ratio of the new sweep span value to the current
sweep span value) used when the MKR ZOOM key is executed as a percentage.
Specify a value between 0.01% and 100% in resolution of 0.01%. Note that the
default value in power-on or preset state is 10%.
ZMAPER <value>
more 2/2
Displays Table B-62, “Marker
®
(Marker
®
/Zoom) menu (1/2),” on page 395.
(N/A)
1. The function to search for a peak is the same as the PEAK key in Table B-66, “Peak (Peak) menu,” on page 398.
Table B-63
Marker
®
(Marker
®
/Zoom) menu (2/2)
Key operation:
[Marker
®
] - more 1/2
Softkey
Function
GPIB command
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......