130
Chapter 6
Calibration
Selecting Appropriate Calibration Method
Selecting Appropriate Calibration Method
The Agilent 4294A has three calibration types: user calibration, port extension
compensation, and fixture compensation. Table 6-1 shows each device connection and its
corresponding adapter selection and calibration. Calibration should be done according to
the particular requirements of each device connection. See “Adapter Setting” on page 67
for adapter selection.
Table 6-1
Device Connection and Corresponding Adapter Selection and Calibration
Device Connection
Adapter
Selection
Accessory
Calibration (
Ö
: required, blank: not required)
User
Calibration
Port Extension
Compensation
Fixture
Compensation
A. Connect a direct connection
type test fixture to the
four-terminal pair port of the
4294A and connect the device
to the fixture.
No adapter
(
NONE
)
16034E
16034H
16044A
16047A
16047C
16047D
16047E
16065A
1
16089A
1
16089B
1
16089C
1
16089D
1
16089E
1
Open/Short
16034G
Open/Short/ Load
(Load comp. is
required when the
frequency is more
than 3 MHz)
16314A
Open/Short/Load
B. Extend the four-terminal
pair port of the 4294A with a
1-meter-long four-terminal pair
configuration and connect the
device.
Four-terminal
pair, 1 meter
(
4TP 1M
)
16048G and test fixture
Same as the fixtures
are directly
connected to the
four-terminal pair
port
16048G and 16452A
Short
16334A
16451B
Open/Short
C. Extend the four-terminal
pair port of the 4294A with a
2-meter-long four-terminal pair
configuration and connect the
device.
Four-terminal
pair, 2 meters
(
4TP 2M
)
16048H and test fixture
Same as the fixtures
are directly
connected to the
four-terminal pair
port
16048H and 16452A
Short
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......