Chapter 6
149
Calibration
User Calibration
6. Ca
libr
ati
on
Step 2.
Press the
USER CAL
key to display the User Calibration Menu.
Step 3.
Press the
DEFINE VALUE
key to display the Calibration Data Definition Menu.
Step 4.
Select one of the calibration data definition keys,
OPEN CONDUCT(G)
,
OPEN CAP(C)
,
SHORT RESIST(R)
,
SHORT INDUCT(L)
,
LOAD RESIST(R)
, or
LOAD INDUCT(L)
, to
confirm or set the calibration data definition.
The current value of the calibration data definition is displayed in the parameter setting
area of the LCD's upper-left side.
Step 5.
Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the calibration data definition.
•
Enter the desired value with the numerical keys (
[0]
through
[9]
,
[.]
,
[-]
) and then press
one of the unit keys (
[G/n]
,
[M/
m
]
,
[k/m]
, or
[
´
1]
).
•
Turn the rotary knob (
m
m
m
m
) until the desired value is set.
•
Press the step keys (
[
][
¯
]
) to set the desired value.
NOTE
When the capacitance value of the calibration data definition is entered with the numerical
keys and one of the unit keys, the unit [fF] is automatically added to the entered value.
Step 6.
Repeat procedures 4 and 5 to set all of the required calibration data definitions.
Calibration Data Definition
Key Stroke
Equivalent parallel conductance value (G) of
OPEN standard used for user calibration
(preset state: 0 S)
OPEN CONDUCT(G)
Equivalent parallel capacitance value (C) of
OPEN standard used for user calibration
(preset state: 0 F)
OPEN CAP(C)
Equivalent series resistance value (R) of
SHORT standard used for user calibration
(preset state: 0
W
)
SHORT RESIST(R)
Equivalent series inductance value (L) of
SHORT standard used for user calibration
(preset state: 0 H)
SHORT INDUCT(L)
Equivalent series resistance value (R) of
LOAD standard used for user calibration
(preset state: 50
W
)
LOAD RESIST(R)
Equivalent series inductance value (L) of
LOAD standard used for user calibration
(preset state: 0 H)
LOAD INDUCT(L)
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......