Chapter 6
137
Calibration
C. Calibration for Four-Terminal Pair, 2-m Extension
6. Ca
libr
ati
on
C. Calibration for Four-Terminal Pair, 2-m Extension
This section describes the calibration procedures to follow when the four-terminal pair port
of the Agilent 4294A is extended with a 2-meter-long four-terminal pair configuration and
the device is connected to the cable’s end with a minimal connection.
Figure 6-3
Calibration for Four-terminal Pair, 2-m Extension
This calibration procedure is applied to the following connections:
•
The 16048H (2-m Cable) is connected to the Agilent 4294A and the device is
connected to the cable’s end with a test fixture.
•
The 16048H (2-m Cable) is connected to the Agilent 4294A and the 16452A (Liquid
Test Fixture) is connected to the end of the cable.
Step 1.
Make sure that the adapter selection is set to “Four-terminal pair, 2-meters (
4TP 2M
).”
See “Adapter Setting” on page 67 for adapter selection.
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......