80
Chapter 4
Preparation of Measurement Accessories
Adapter Setting
NOTE
When the load data measurement is completed, the softkey label changes to
OPEN [DONE]
.
Step 8.
Connect the SHORT (0
W
, HRM 504) to the 3.5-mm port (the SHORT is furnished with the
42941A Impedance Probe).
Step 9.
Press the
SHORT [-]
key to start the short data measurement.
NOTE
When the short data measurement is completed, the softkey label changes to
SHORT [DONE]
.
Step 10.
Remove the SHORT (0
W
, HRM 504) from the 3.5-mm port of the 42941A. Then connect
the LOAD (50
W
, HRM 601) to the 3.5-mm port (the LOAD is furnished with the 42941A
Impedance Probe).
Step 11.
Press the
LOAD [-]
key to start the load data measurement.
NOTE
When the load data measurement is completed, the softkey label changes to
LOAD [DONE]
.
Step 12.
Press the
done
key to start calculating the adapter setup data from the measured phase
compensation, OPEN, SHORT, and LOAD data. The adapter setup data is automatically
saved to the EEPROM (nonvolatile memory) in the Agilent 4294A.
NOTE
Press the
cancel
key to cancel the adapter setup process. If the adapter setup process is
canceled, the previously obtained data is saved as the available adapter setup data.
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......