328
Chapter 10
Specifications and Supplemental Performance Characteristics
Basic Characteristics
Figure 10-1
dc Bias Range (SPC)
Sweep Characteristics
Available sweep parameters
Frequency, signal voltage, signal current, dc bias
voltage, dc bias current
Sweep types
Linear, log
1
, list
1
, zero-span, manual
Manual sweep
Available for all sweep types
Sweep directions
Up-sweep, down-sweep
Number of measurement
points
2 to 801
List Sweep
Available setup
parameters for each segment
Sweep frequency range, number of measurement
points, signal level (voltage or current), dc bias
(voltage or current), measurement bandwidth, point
averaging factor
Number of segments
1 to 18
Sweep span types
Segment span, single span
Delay time
Types
Point delay, sweep delay
Range
0 to 30 sec
Resolution
1 msec
1. Frequency sweep only.
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......