162
Chapter 7
Setting Up the Display of Measurement Results
Selecting the Graph Axis Format
When Using Complex Parameters (COMPLEX Z-Y)
When the Agilent 4294A is configured to perform a COMPLEX Z-Y measurement
(complex impedance for Trace A, complex admittance for Trace B), use the following
procedure to select the graph axis formats:
Step 1.
Activate the trace for which you want to set the axis format.
Step 2.
Press the
[Format]
key to display the Format menu.
Step 3.
Select the desired graph axis format by pressing the appropriate key:
Figure 7-3
Complex plane whose Y axis indicates complex admittance (with only Trace B
displayed)
To activate:
Press:
Trace A
[A]
Trace B
[B]
To select:
Press:
Complex plane (where the horizontal and
vertical axes use real and complex values,
respectively)
COMPLEX
Polar chart
POLAR
Содержание 4294A
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...8 ...
Страница 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Страница 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Страница 100: ...92 Chapter5 Setting Measurement Conditions Setting Sweep Range Figure 5 7 After MKR CENTER Figure 5 8 ...
Страница 101: ...Chapter 5 93 Setting Measurement Conditions Setting Sweep Range 5 Setting Measurement Conditions Figure 5 9 ...
Страница 106: ...98 Chapter5 Setting Measurement Conditions Setting Number of Points NOP Figure 5 13 Number of points 801 ...
Страница 330: ...322 Chapter9 Setting Using Control and Management Functions Performing Self Diagnosis of the Agilent 4294A ...
Страница 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Страница 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Страница 462: ...454 AppendixD Initial Settings Initial Settings Settings that can be Saved Recalled Settings that can be Backed Up ...
Страница 468: ......