VERIFICATION PROCEDURESVP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
3
5. If check fails, adjust Vent Wash Block.
a. Perform
“VP - 3.8 Rack to Laser Shield Check” (Page 3-39)
.
b. Loosen Retaining Nut (
).
c. Move Vent Wash Block to align needle in Tool.
d. Ensure flat side of Wash Block is parallel to Contact Sensor Board.
Figure 3-71 Vent Needle Alignment
--- END ---
Move Vent
Wash Block to
Adjust Needle
Loosen
Retaining
Nut
SL-182A