28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 13.1 Lead Bubble Check/Adjustment
3 - 155
3
d. Aspirate remaining cleaner by manually opening Solenoid 96 and release lever from Sample
Peristaltic Pump P2.
Cleaner moves through Shear Valve via Peristaltic Pump to Waste Chamber 2.
NOTE:The typical clot is in the T-fitting that joins the peristaltic pump to the Shear Valve and
Solenoid 92.
e. Run background counts to flush cleaner.
5. Load 3 Vacutainer
®
tubes (G Loader - use Sarstedt tube), each with 1-3 ml of control or blood
sample.
6. Enter Normal Mode.
7. Press START.