VP - 3.3 Platen Wall Adjustment
Purpose: Perform this procedure to adjust Platen Walls, Rack Guides, and Beam Block.
1. Determine if Platen is welded or non-welded. Non-welded Platen has 14 wall screws on top of
Platen.
2. If Platen is non-welded, go to Step 3. If Platen is welded, go to Step 4.
3. Align Non-Welded Platen:
a. Loosen Platen Wall screws (1/2 turn) at right end of Platen.
b. Insert Rack Guide Alignment Tool at Right end (1) of Platen. Ensure Walls are tight against
Tool then tighten screws.
Special Tools
DESCRIPTION
QTY
PN
Rack Guide
Alignment Tool
1
9420084