CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 109
3
F Loader
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Clear all racks from under mixing paddle.
3. Press (Z
↓
) key down until mixing paddle stops.
4. Check that distance between platen to knurled edge of mixing paddle is 3.25 in. (82.5mm) (