3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.3 Mixing Home Sensor Check/Adjustment
3 - 89
3
VP - 6.3 Mixing Home Sensor Check/Adjustment
Purpose: Perform this procedure to verify and adjust Mixing Head Assembly Home Position.
All Loaders
1. Clear Racks from under Mixing Paddle.
2. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
3. Press (
↓
Z
↑
) keys to move the Mixing Paddle all the way down then all the way up and check
measurement (
).
NOTE:Measure from knurled edge of Mix Paddle not rounded tip.
4. If check fails, adjust Home Sensor (
).
NOTE:If check continues to fail, ensure Shims (2) are present below Mixing Assembly. See
5. Go to
“VP - 6.4 Mixing Contact Sensor Check/Adjustment” (Page 3-91)
Special Tools
DESCRIPTION
QTY
PN
None