CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.6 X-Axis Solenoid to Guide Rod Check
3 - 31
3
VP - 3.6 X-Axis Solenoid to Guide Rod Check
Purpose: Perform this procedure to verify that the X-Axis Solenoid Slide is square (perpendicular)
to the X-Axis Guide Rods,
. Required for [NON-CE Mark] Loaders only.
1. Locate X-Axis Home Sensor Flag below Platen, attached to X-Axis Solenoid Slide (
).
2. Verify color of X-Axis Sensor Flag (
GRAY [CE MARK] - No adjustment required. Install Platen (if removed). Go to
Laser Shield Check” (Page 3-39)
BLACK [NON-CE MARK] - Adjustment required. If Platen is removed, do not install. Go to Step
3.
3. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
4. If Platen is installed, go to Step 5, if Platen is Not installed, go to Step 6.
Special Tools
DESCRIPTION
QTY
PN
Rack Guide
Alignment Tool
1
9420084
Feeler Gauge Set
1
1-43627-01