28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 15.1 Barcode Reader Configuration (Europe Only) 3 - 165
3
a. Press [F5].
Main Menu displays MAIN ---> COMMUNICATIONS message at bottom of screen.
b. Press CapsLock key.
CapsLock is activated.
NOTE:VBCDIAG requires commands in capital letters.
c. Press [SPACE BAR] two (2) times.
MAIN ---> CODE TYPES displays at bottom of screen.
d. Press [ENTER].
CODE TYPES screen displays with CODE TYPES ---> NARROW MARGINS - DISABLED
message (bottom of screen).
e. Press [SPACE BAR] until desired Barcode type displays (bottom of screen).
f.
Press [ENTER].
Desired Code Type displays (bottom of screen) with Parameter available for editing.
g. Press [SPACE BAR] until desired Parameter displays (bottom of screen),
h. To edit ENABLE/DISABLE fields:
-
Press [SPACE BAR] to toggle between ENABLE/DISABLE.
-
Press [ENTER].
i.
To edit CODE LENGTH:
-
Type Code Length value.
-
Press [ENTER].
-
Type Code Length value again.