CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.12 Platen to Laser Shield Check
3 - 47
3
VP - 3.12 Platen to Laser Shield Check
Purpose: Perform this procedure to verify Platen is aligned (front to back) with the Tower.
1. Ensure Laser Shield hardware is tight.
2. Verify Laser Shield is slightly recessed from Platen Wall. Check Left and Right end of Laser
Shield (
).
• IF Check is OK, go to
“VP - 3.14 Pusher Arm to Platen Check” (Page 3-51)
• IF Check Fails, go to
“VP - 3.13 Platen to Laser Shield Adjustment” (Page 3-49)
.
Special Tools
DESCRIPTION
QTY
PN
None