28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.14 Pusher Arm to Platen Check
3 - 51
3
VP - 3.14 Pusher Arm to Platen Check
Purpose: Perform this procedure to verify the X-Axis Pusher Arms are aligned in the Platen Wall
Cutout,
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Press Y key on Loader Keyboard until Pusher Arms are retracted.
3. Pusher head must be flush or, in the case of E and F Loaders, recessed approximately 1/2 of
platen wall thickness (
).
4. Pusher head should be approximately centered in platen wall cutout (
).
• If Check is OK, go to
“VP - 3.16 X-Axis Solenoid to Rack Guide Check” (Page 3-55)
.
• If Check Fails, go to
“VP - 3.15 Pusher Arm to Platen Adjustment” (Page 3-53)
.
Special Tools
DESCRIPTION
QTY
PN
None