VERIFICATION PROCEDURESVP - 12.1 Blood In-Line Sensor Check/Adjustment
3 - 152
3
8. Run the 0.80 sample or inject sample into line with syringe. With Blood in the In-Line Sensor,
TP4 (+) = +2.4 to +2.6VDC.
Figure 3-81 Meter to TP4 and TP1
9. Connect DC meter to Sensor board TP1 (+) to TP3 (-) GND.
10.With Blood in the In-Line Sensor, TP1 (+) = +0.3 to +3.25VDC.
Blood in
Sensor Path
TP3 = GND
TP4 = +2.5VDC
@0.8 M/
µ
l Dilution
TP1=+0.3 to 3.25VDC
@0.8 M/
µ
l Dilution
SL-107C
Test pins on opposite
side of board