VERIFICATION PROCEDURESVP - 3.11 X-Axis Solenoids to Platen Channel Adjustment 3 -
3
VP - 3.11 X-Axis Solenoids to Platen Channel Adjustment
Purpose: Perform this procedure to align the Platen Channels with the X-Axis Guide Rods.
1. Perform Check procedure
“VP - 3.10 X-Axis Solenoids to Platen Channel Check” (Page 3-43)
2. Loosen (4) Platen Mounting screws 1/2 turn (
).
NOTE:Though the location of the Platen Screws may be different, there is a slot in the platen for
each Platen Screw. Use this slot as a means of locating the Platen Screws.
3. Adjust Platen until solenoid travel is parallel to Platen Channel and gap is same for both
Solenoids (
).
4. Verify rack to Laser shield check is still OK. Perform
“VP - 3.8 Rack to Laser Shield Check” (Page
,
5. Go to
“VP - 3.12 Platen to Laser Shield Check” (Page 3-47)