CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 1 Power Supply Check [NON-CE MARK]
3 - 10
3
4. Verify the following:
• LED on +24V Power Supply ON.
• LED on +5V Power Supply ON (E and G Loaders - View from Front; F Loader - View from
Rear).
• LED on Blood In-Line Sensor ON if no blood in Sample Line.
• F Loader LED ON (located on Barcode Power Supply).
-
DS3 LED ON (located on Barcode Decoder Board).
5. Verify voltage at CPU Main Board, U8 Pin 3 = +15±0.8V.
6. Verify voltage at ±12V Power Supply equals 12V±0.4 (