CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 10.2 Tube Positioning Arm Alignment
3 - 135
3
VP - 10.2 Tube Positioning Arm Alignment
Purpose: Perform this procedure to adjust Tube Positioning Arm Length, Energized and
Non-Energized.
NOTE:E and F Loaders Only - G Loader does not have a Tube Positioning Arm.
1. Perform Check procedure
“VP - 10.1 Tube Positioning Arm Check” (Page 3-133)
2. Remove Sample Loader Covers and Racks.
3. Disconnect tubing between analyzer and Sample Loader. Do Not bend or kink sample tubing.
4. Turn Sample Loader to access rear.
5. Ensure tube positioning arm solenoid is straight.
F Loader [Only OLD With Clamp] - (Others not Adjustable)
a. Loosen bracket and adjust as necessary (
Special Tools
DESCRIPTION
QTY
PN
None