28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURES VP - 11.1 Barcode Check
3 - 141
3
VP - 11.1 Barcode Check
Purpose: Perform this procedure to verify Barcode Reader Operation.
1. Perform Platen checks
2. Clean Laser shield glass front and back, if old style Laser Shield (without detachable Window
Mount), ensure glass extends to bottom of notch in platen wall (
). If glass needs
adjustment, go to
“VP - 11.2 Barcode Adjustment” (Page 3-145)
a. Remove shield and clean glass on both sides.
b. Inspect and clean Barcode Reader lenses.
c. Install laser shield.
3. Allow system to place tool under paddle.
a. Block safety cover sensor.
b. Ensure Data Station is on, system is initialized, and Closed Sampler is selected.
c. Load Needle Alignment Tool in tube position number 1 of a rack and install rack in #1 position
).
d. Load other nine (9) racks.
e. Put Loader in Normal Mode.
f.
Press START on Sample Loader.
Special Tools
DESCRIPTION
QTY
PN
Needle Alignment
Tool
1
9420081