28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 93
3
5. Verify Contact Sensor Board Position.
E and F Loaders
[CE MARK] Top of Mixing Paddle must be .040-.060 in. (1-1.5 mm) below bottom of Sensor
when Mixing Head is in Home Position (Z
↑
). See
.
Figure 3-46
[NON-CE MARK] E Loader - Top of Mixing Paddle must be flush with bottom of Sensor
(when Mixing Assembly is home,
.040 - .060 in
(1.0 - 1.5 mm)
[CE MARK]
SL-269
E- Loader