28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
3
VP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
Purpose: Perform this procedure to verify and adjust Vent Needle Alignment, F Loader Only,
.
1. Perform Steps 1, 2, and 3 of
“VP - 9.1 Sample Wash Block Check/Adjustment” (Page 3-125)
.
2. Ensure proper Needle is installed and Needle Port is centered on Wash Port (
).
Special Tools
DESCRIPTION
QTY
PN
Needle Alignment
Tool
1
9420081
Rack Guide
Alignment Tool
1
9420084