background image

 

                                                         

           

KJ6013 

Digital Capacitance Meter 

 

 

 

Operation Manual 

 

 

 

 

 

 

Содержание KJ6013

Страница 1: ...KJ6013 Digital Capacitance Meter Operation Manual ...

Страница 2: ...tment manual adjusting range is 20pF Display Large size LCD display Max 1999 counts Range 9 ranges measuring range is from 0 1pF to 20mF Sampling time 0 5 seconds Calibrating adjustment a panel zero adjuster for zeroing adjustment Over range indication 1 display Low battery indicator display Operating humidity 80 RH Operating temperature 0 40 32 104 Storage temperature 10 50 14 122 Power supply si...

Страница 3: ...ency 200pF 199 9pF 0 1pF 800Hz 2nF 1 999nF 1pF 800Hz 20nF 19 99nF 10pF 800Hz 200nF 199 9nF 100pF 800Hz 2µF 1 999µF 1000pF 800Hz 20µF 19 99µF 0 01µF 80Hz 200µF 199 9µF 0 1µF 8Hz 2mF 1 999mF 1µF 8Hz 20mF 19 99mF 10µF 8Hz Unit conversion formula 1Farad 10 3 Millifarads 10 6 Microfarads 10 9 Nanofarads 10 12 Picofarads Accuracyis pencentage of reading number of last digits at 23 5 80 RH 200pF 0 5 1 0 ...

Страница 4: ...and terminal it will cause the battery dissipation rise This C meter is intended for measuring the capacitance value of capacitor It is not intended for determining the Q factor for above reactive components Misleading readings may be obtained if the measurement of capacitance of a resistor is attempted When measuring in circuit that circuit must be switched off and de energized before connecting ...

Страница 5: ...e measured Confirm display show 0000 firstly low test range such as 200 pF 2 nF 20 nF need to adjust ADJ to be zero Connect the measured capacitor and press TEST DIS button to discharge observe polarity when connecting polarity capacitor Release TEST DIS button measuring Read the display the value is direct reading in the unit pF nF uF or mF indicated at the selected range switch If display show 1...

Страница 6: ...tioned test leads remember that the leads maybe introduce a measurable capacitance to influence the accuracy of measurement As a tip of measurement the test lead capacitance may be measured by opening the leads and recording it down then subtracting it to obtain a more accurate measuring result e Capacitors especially electrolytic it often have a wide tolerance Do not be surprised if the measured ...

Страница 7: ...ean the surface of the instrument Take out of the battery to prevent the battery electrolyte leakage if don t use for long time Battery replacement Replace the battery immediately when LCD indicates symbol steps as the following Take out of the screw and remove battery cover Remove 9V battery and replace with another new battery with same type or equivalent Reinstall the battery cover and fasten t...

Страница 8: ...7 new one with same type or equivalent Reinstall the bottom case and fasten the screws End This operating manual subject to change without notice ...