CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 5.1 Tube Detect Sensor (Beam Block) Check
3 - 67
3
2. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
3. Set BAR (1/0) Switch to 1.
Figure 3-31 White Beam Block
4. Put 5 Racks on right side of Platen.
Detector
LED
SL-225