28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 5.2 Tube Detect Sensor (Beam Block) Adjustment
3 - 70
3
VP - 5.2 Tube Detect Sensor (Beam Block) Adjustment
Purpose: Perform this procedure to align Tube Detector (Beam Block).
1. Perform Check procedure
“VP - 5.1 Tube Detect Sensor (Beam Block) Check” (Page 3-66)
2. Remove Covers and, if needed, move Keyboard to Access CPU Main Board.
3. For Black Beam Block, go to Step 4, for White Beam Block, go to Step 5.
Special Tools
DESCRIPTION
QTY
PN
Resistor 3.92k
Ω
([NON-CE MARK]
only)
1
4703062