28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.3 Platen Wall Adjustment
3 - 24
3
4. Align Welded Platen:
a. Welded Platen walls cannot be straightened. If walls are not square, replace Platen.
b. It may be necessary to adjust Rack Guide and/ or Beam Block for Tool to fit properly.
To Adjust Rack Guide:
-
Loosen (2) screws attaching Rack Guide to Platen.
-
Insert Rack Guide Alignment Tool, adjust Rack Guide, and tighten hardware.
To Adjust Beam Block:
-
Go to
“VP - 3.5 Rack Clearance Adjustment” (Page 3-28)
.
5. Go to