CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 14.1 Y-Axis Operation Check/Adjustment
3 - 160
3
b. Press Y again. Pusher Arm should retract immediately and not extend farther into Platen.
5. If Step 3 check fails, perform the following steps:
a. Remove Platen.
b. Loosen 2 set screws at left end of shaft (
).
c. Adjust Y-Axis Drive Mechanism.
d. Pull both Push Arm slides away from platen.
e. Verify Connecting Bars are Horizontal.
f.
Verify position of Sensor Flags.
-
Ensure slot in Home Sensor Flag is positioned properly (see
).
g. Tighten set screws on left end of shaft.
h. Check Drive Belt tension.
i.
Tighten all Y-Axis hardware.
j.
Install Platen and perform all Platen checks (