CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.5 Mixing Limit Sensor Check/Adjustment
3 - 111
3
G Loader
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Clear all racks from under mixing paddle.
3. Press (Z
↓
) key down until mixing paddle stops.
4. Plate behind mixing paddle and Sensor Mounting Bracket should clear the laser shield and the
vent wash block by
≈
.020 in. (0.5mm) (
5. If check fails, adjust Limit Sensor.