CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 101
3
g. Verify REPEAT LED Flashers.
CAUTION:Ensure slot in Paddle faces Sensor.
h. Verify Contact Sensor turns Off.
-
[CE MARK] Release the Mixing Paddle. Verify the Sensor is unblocked and REPEAT LED
is Off.
-
[NON-CE MARK] Position Mixing Paddle to block 1/2 of Sensor. Verify REPEAT LED is
Off.
9. If Check Fails:
[CE MARK] adjust R7 (with Mixing Paddle positioned as indicated in
[NON-CE MARK] adjust VR6 (DO NOT exceed position shown in
NOTE:If unable to perform adjustment, add reflective tape or white correction fluid to top of
Mixing Paddle and readjust.
10.Go to
“VP - 6.5 Mixing Limit Sensor Check/Adjustment” (Page 3-103)