CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 7.2 Mixing Paddle Overcurrent Alarm Adjustment 3 - 119
3
f.
Connect Mixing Motor Current Tool to Mixing Motor #2 (
g. Press R. Loader should beep.
h. If no beep, turn R - 84 Clockwise until Loader beeps.
i.
Reconnect Mix Motor # 2.
j.
Go to Step 11 (Page
). Procedure is complete.