CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.6 X-Axis Solenoid to Guide Rod Check
3 - 35
3
c. Check distance between Tool and solenoids using a Feeler Gauge. Distances should be the
same (±.005 in., 0.127mm).
d. If Check is OK, Install Platen then go to
“VP - 3.8 Rack to Laser Shield Check” (Page 3-39)
e. If Check Fails, go to
“VP - 3.7 X-Axis Solenoid to Guide Rod Adjustment” (Page 3-36)
.
.
Figure 3-14 Solenoid Parallelism
--- END ---
Same
Distance
Same
Distance
SL-235
Solenoid
Alignment Tool